TEL: +86 571 56623320 EMAIL: [email protected]
Summary:
Purpose
Our Quick Step Human Neuregulin 4,NRG-4 ELISA Kit is to assay NRG-4 levels in Human serum, plasma, culture media or any biological fluid.
Principle
This ELISA kit uses Sandwich-ELISA as the method. The Microelisa stripplate provided in this kit has been pre-coated with an antibody specific to NRG-4. Standards or samples are added to the appropriate Microelisa stripplate wells and combined to the specific antibody. Then a Horseradish Peroxidase (HRP)- conjugated antibody specific for NRG-4 is added to each Microelisa stripplate well and incubated. Free components are washed away. The TMB substrate solution is added to each well. Only those wells that contain NRG-4 and HRP conjugated NRG-4 antibody will appear blue in color and then turn yellow after the addition of the stop solution. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of NRG-4. You can calculate the concentration of NRG-4 in the samples by comparing the OD of the samples to the standard curve.
Typical Data
The standard curve of QS3221Hu is provided for demonstration only. A standard curve should be generated for each set of samples assayed.
Standard
Concentration
OD Value
Average OD Value
Blank Well
0pg/ml
0.046
0.0435
0.041
S1
24pg/ml
1.989
1.9925
1.996
S2
16pg/ml
1.331
1.3420
1.353
S3
8pg/ml
0.694
0.7025
0.711
S4
4pg/ml
0.391
0.3865
0.382
S5
2pg/ml
0.194
0.1975
0.201
Scan Wechat Qrcode
Scan Whatsapp Qrcode