TEL: +86 571 56623320 EMAIL: [email protected]
Summary:
Purpose
Our Quick Step Mouse Excitatory Amino Acid Transporter 2 (EAAT2/SLC1A1/GLT-1)ELISA Kit is to assay EAAT2/SLC1A1/GLT-1 levels in Mouse serum, plasma, culture media or any biological fluid.
Principle
This ELISA kit uses Sandwich-ELISA as the method. The Microelisa stripplate provided in this kit has been pre-coated with an antibody specific to EAAT2/SLC1A1/GLT-1. Standards or samples are added to the appropriate Microelisa stripplate wells and combined to the specific antibody. Then a Horseradish Peroxidase (HRP)-conjugated antibody specific for EAAT2/SLC1A1/GLT-1 is added to each Microelisa stripplate well and incubated. Free components are washed away. The TMB substrate solution is added to each well. Only those wells that contain EAAT2/SLC1A1/GLT-1 and HRP conjugated EAAT2/SLC1A1/GLT-1 antibody will appear blue in color and then turn yellow after the addition of the stop solution. The optical density (OD) is measured spectrophotomeEAAT2/SLC1A1/GLT-1ally at a wavelength of 450 nm. The OD value is proportional to the concentration of EAAT2/SLC1A1/GLT-1. You can calculate the concentration of EAAT2/SLC1A1/GLT-1 in the samples by comparing the OD of the samples to the standard curve.
Typical Data
The standard curve of QS1060Mo is provided for demonstration only. A standard curve should be generated for each set of samples assayed.
Standard
Concentration
OD Value
Average OD Value
Blank Well
0pg/ml
0.052
0.0460
0.040
S1
3600pg/ml
1.912
1.9075
1.903
S2
2400pg/ml
1.388
1.3840
1.380
S3
1200pg/ml
0.708
0.7100
0.712
S4
600pg/ml
0.403
0.4000
0.397
S5
300pg/ml
0.189
0.1960
0.203